51
1000.0
0.7
1.1
1.5
3.7
4.1
4.6
5.0
6.7
7.4
1100.0
0.9
1.3
1.8
4.3
4.7
5.3
5.7
7.5
8.2
1200.0
1.0
1.4
2.0
4.9
5.4
6.0
6.4
8.4
9.1
1300.0
1.1
1.6
2.3
5.5
6.0
6.7
7.2
9.2
10.0
1400.0
1.3
1.8
2.6
6.2
6.7
7.4
7.9
10.1
10.9
1500.0
1.4
2.0
2.8
6.8
7.4
8.1
8.7
11.0
11.9
1600.0
1.6
2.3
3.1
7.5
8.1
8.9
9.5
11.9
12.7
1700.0
1.8
2.5
3.5
8.2
8.9
9.7
10.3
12.8
13.7
1800.0
1.9
2.8
3.8
8.9
9.6
10.5
11.1
13.7
14.6
1900.0
2.1
3.0
4.2
9.6
10.4
11.3
12.0
14.6
15.5
2000.0
2.3
3.3
4.5
10.4
11.2
12.1
12.8
15.5
16.5
2100.0
2.5
3.6
4.9
11.2
12.0
12.9
13.6
16.4
17.4
2200.0
2.8
3.9
5.3
12.0
12.7
13.8
14.5
17.3
18.3
2300.0
3.0
4.2
5.7
12.7
13.6
14.6
15.3
18.3
19.3
2400.0
3.2
4.5
6.1
13.5
14.4
15.5
16.2
19.2
20.3
2500.0
3.5
4.9
6.6
14.3
15.2
16.3
17.1
20.2
21.2
2600.0
3.7
5.2
7.0
15.1
16.1
17.2
18.0
21.1
22.2
2700.0
4.0
5.6
7.5
15.9
16.9
18.1
18.9
22.1
23.1
2800.0
4.2
5.9
8.0
16.8
17.8
19.0
19.8
23.0
24.1
2900.0
4.5
6.3
8.4
17.6
18.6
19.9
20.7
24.0
25.1
3000.0
4.8
6.7
8.9
18.4
19.5
20.7
21.6
24.9
26.0
Example: For measured target temperature of 160 deg C, M67 spectral band code of B and uncertainty of emissivity setting of
5%, the uncertainty of temperature measurement is 5.0 deg C