M310 Manual Communications Port 19
Controller response:
%0101R0902.000N8<CR>
Command line to change blackbody setpoint to 30C:
$0101W0930.000F5<CR>
Controller response:
%0101w100K2<CR>
Note that bold characters above represent the Checksum Actual characters may be different
due to the calculation of checksum. Refer to the Appendices for a subroutine that calculates
checksum for each writing string. Also note that checksum must be used in all communications.
In the Appendices is a typical QBASIC program that reads current temperature and increments
setpoint at 10 °C intervals: