Lumasense-technologies 3434i SF6 Leak Detector Manual de usuario Pagina 19

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 191
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 18
Chapter 2
______________________________________________________________________
_____________________________________________________________________________
BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S
Page 19 of 191
2.4.3 Checking/Changing the Text line Terminator.
The Text line Terminator must be set using the push-buttons on the
front of the monitor.
Press SET-UP S2 S3
Press SET-UP to exit the set-up mode
2.4.4 Checking/Changing the Data Log and Error Log
Options.
The Data Log and Error Log must be set using the push-buttons on
the front of the monitor.
Press SET-UP S3 S1 S1 S3 S1 S1
Press SET-UP to exit the set-up mode
2.5 Setting the Ethernet Communication Parameters
The LumaSoft Gas 7810 application with Ethernet communication is
able to communicate with the 3434i gas monitor through a local
Ethernet network.
The homepage of the 3434i gas monitor can also be displayed in a
web browser, when the 3434i gas monitor is connected to a local
Ethernet network.
2.5.1 Checking/Changing the Ethernet Communication
Parameters
The communication parameters for the Ethernet interface can be set
by using the push-buttons on the front of the monitor.
1. Press SET-UP S3 S1 S3 S2. The screen display now shows the
following text.
2. Press S1 to enter the Ethernet setup. During this setup you can
also use the andkeys to go back and forth between the
SELECT ADDRESSABLE SET-UP BRANCH
Ethernet IEEE 488
Vista de pagina 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 190 191

Comentarios a estos manuales

Sin comentarios