4
2 Technical data
Calibration source:
Temperature: depending on type: 45°C, 70°C, 75°C, 95°C, 100°C, 110°C,
120°C, 140°C or 150°C
Emissivity : 0.98 ±0.004 (for = 2 ... 5.4 µm and = 8 ... 14 µm)
Aperture diameter: 50.8 mm
Warm-up time: max. 5 (IRC 45) up to 15 minutes (IRC 150)
Temperature uncertainty: 0.4°C for T
amb
= 10...40°C (IRC 45-120)
0.6°C for T
amb
= 0...10°C (IRC 45-120)
0.5°C for T
amb
= 10...40°C (IRC 140-150)
0.7°C for T
amb
= 0...10°C (IRC 140-150)
Repeatability: 0.2°C
Stability: 0.1°C
Temperature uniformity: 0.2°C (central area Ø 45 mm)
Operating temperature: 0 ... 40°C, temporary (2 min) up to 70°C
Storage temperature: 0 ... 70°C
Relative humidity: 10 ... 85%, no condensing conditions
Status LED: green: warm-up
orange: ready for operation
red: overheating
Power supply: 24 V DC, max. 1 A
Protection class: IP50 (EN 60529)
Weight: 0.85 kg
Dimensions: 64.5 x 81.0 x 133.5 (Ø x D x H)
CE sign: according to EU directives about electromagnetic immunity
Power supply:
Power supply: 230 V AC, 50 Hz
Output: 24 V DC, 1 A
Protection class: IP50 (EN 60529)
Weight: 1.15 kg
CE sign: according to EU directives about electromagnetic immunity
2.1 Dimensions
LED
Connection
24 V supply
Ø 50.
8
Ø 64.5
Ø 12
Ø 49.3
81
.
0
25.5
69
1
0
1.2
10