BoilerSpection IM Manual Troubleshooting 27
6 Troubleshooting
6.1 Lens
Symptom Probable Cause Corrective Action
Poor image – miss
slag buildup
Lens surface contamination due to dust/dirt in
the boiler environment
Clean lens with q-tip or install
air purge
Lens surface contamination during
installation/maintenance
Clean lens
Image offset caused by excessive heat Install air cooling
Focus change caused by excessive heat
Excessive heat causes lens component to
become opaque
Contact LumaSense
(refer to section 1.7)
Loss of image Filters cracked because of thermal cycling
Contact LumaSense
(refer to section 1.7)
Nose tip damaged because of excessive heat
Nose tip damage because of vibration failure
Damage to mechanical/internal fixtures
caused by excessive heat
Cooling shroud damaged caused by
missalignment
6.2 Camera
Symptom Probable Cause Corrective Action
Complete Power
Failure
Vibrations/internal failures Contact LumaSense
(refer to section 1.7)
Loose power connections Disconnect battery and check
all power wiring connections
Excessive heating Air cooling needed or add
heat shields if hot air or
radiation from the furnace is
directly heating the failed
component or enclosure
Battery Discharged Recharge Battery