MANUA LM390Blackbody Calibration Source
M390 Manual Introduction 102.2 Technical DataThisModelModel Part No. Exit Aperture Temperature Range Power[ ] M390-A1 14029-A1 0.625 in 600-
M390 Manual Installation 113 Installation3.1 Installing the M390To install the M390:1. Allow the M390 to warm-up to room temperature for at least 2
M390 Manual Installation 12To ensure consistent document formatting, this page was intentionally left blank.
M390 Manual Operation 134 Operation4.1 Starting UpTo start up the M390:1. Turn on the power switch/breaker and allow 20 seconds for the system to s
M390 Manual Operation 14Eurotherm controller4.2.1 Other Controller ModesIf, for some reason, the power must be turned off while monitoring temperat
M390 Manual Operation 158. Press the UP button to set new units °F (or K for Kelvin, if desired). Confirm the changewith the RETURN button or press
M390 Manual Operation 164.5.1 RS485 OptionRS485 two wire is available as an option. The connection scheme is shown below:RS485 for multi-drop appl
M390 Manual Operation 174.6 PID Tuning ConstantsThe PID tuning constants have been factory set for optimum performance and are noted below.Heating
M390 Manual Operation 18Typical Radiance temperature calibration look-up table (LUT) used in our blackbodyTo gain access to the custom look-up tabl
M390 Manual Operation 19Note: Calibration points do not have to be used as pre-programmed. To dial in adifferent calibration point, simply change a
Confidential InformationThe material contained herein consists of information that is the property of LumaSenseTechnologies and intended solely for us
M390 Manual Operation 20To ensure consistent document formatting, this page was intentionally left blank.
M390 Manual Maintenance 215 Maintenance5.1 Cavity LifeThe life expectancy of the graphite cavity depends on the time and temperature of operation.T
M390 Manual Maintenance 225.4 Furnace Core DisassemblyThe graphite core can be replaced without disconnecting cooling or purge tubing. Refer toexpl
M390 Manual Maintenance 235.5 Fuse LocationsThe M390 has several fuses. Refer to the table below. See the appendix for photo-view of fuselocations
M390 Manual Maintenance 245.6 Recommended Spare PartsModelTemperatureRange Cavity LengthReplacementCavity KitQuartz Tube, ElectrodeSeats, and O-Rin
M390 Manual Troubleshooting 256 TroubleshootingSymptom Tests/ Solutions1. Blackbody Does not heat up Is there both Argon purge gas and coolingwater
M390 Manual Troubleshooting 269. Blackbody heats up and is glowing, butthe controller does not indicate temperature> beginning of range.Check th
M390 Manual Appendix A: Drawings 27Appendix A: Drawings
M390 Manual Appendix A: Drawings 28
M390 Manual Appendix B: ASCII/BASIC 29Appendix B: ASCII/BASICTypical ASCII request string to read the blackbody temperature:'-------make a req
M390 Manual Contents iiiContents1 General Information ...
M390 Manual Appendix B: ASCII/BASIC 30A simple BASIC program to send blackbody setpoint:CLS'-----open comm port 1 for 9600 baud, 7,E,1-----OPE
M390 Manual Appendix C: Declaration of Conformance 31Appendix C: Declaration of Conformance
M390 Manual Appendix C: Declaration of Conformance 32To ensure consistent document formatting, this page was intentionally left blank.
M390 Manual Index 33IndexCCavity life 21Cleaning 24Controller Calibration 17Controller Modes 14DDisposal 7EEurotherm iTools Software 17FFurnace Cor
M390 Manual Index 34To ensure consistent document formatting, this page was intentionally left blank.
M390 Manual Index 35
M390 Manual Contents iv6 Troubleshooting ...
M390 Manual General Information 51 General Information1.1 Information about the user manualCongratulations on choosing the high quality and highly
M390 Manual General Information 6Warning: DO NOT USE NITROGEN (N2) FOR PURGE GAS. AT HIGH TEMPERATURES,CYNIDE GAS COULD BE GENERATED.1.3 Limit of L