IMPAC pyrometers IN 210 · IN 210/4 · IN 210/5
4
1 Technical data
Temperature range: IN 210: -32 ... 900°C (-25.6 ... 1652°F)
IN 210/4: 300 ... 1000°C (572 ... 1832°F)
IN 210/5: 100 ... 1200°C (212 ... 2192°F)
Sub range: User adjustable within the temperature range
(via service interface). Minimum span 51°C
Data handling: Digital
Spectral range: IN 200: 8 ... 14 µm
IN 200/4: 3.9 µm
IN 200/5: 5.14 µm
IR detector: Thermopile
Power supply: 24 V DC ± 25%, ripple < 0.5 V
Power consumption: max. 0.6 W
Analog output: Linear current 4 to 20 mA
Note: At ambient temperatures > 75°C the output signal is set to
approx. 3.9 mA
Load:
max. 700 at 24 V power supply; 400 (18 V); 1000 (30 V)
Service interface: for parameter programming with portable battery driven setup
device HT 6000
Emissivity ε: 20 to 100% adjustable via service interface
Response time t
90
: 120 ms; up to 10 s adjustable via service interface
Uncertainty: 1% of measured value in °C + 1°C (ε = 1, t
90
= 1 s, T
amb.
= 23°C)
Repeatability: 0.5% of measured value in °C + 1°C (ε = 1, t
90
= 1 s, T
amb.
= 23°C)
Noise Equivalent Tem-
perature Difference
(NETD), ( = 1, ε = 1):
IN 210: with t
90
= 120 ms: 0.2°C (23°C measured temp.)
with t
90
= 1 s: 0.05°C (23°C measured temp.)
IN 210/4: with t
90
= 120 ms: 0.6°C (500°C measured temp.)
with t
90
= 120 ms: 0.2°C (1100°C measured temp.)
IN 210/5: with t
90
= 120 ms: 0.15°C (300°C measured temp.)
with t
90
= 120 ms: 0.1°C (1100°C measured temp.)
Ambient temperature: 0 to 70°C
Storage temperature: -20 to 70 °C
Weight: approx. 450 g
Housing: Stainless steel
Operating position: Any
Protection class: IP65 (DIN 40 055)
CE-label: according to EU directives about electromagnetic immunity