IN 6/78-L Operating Manual Data format UPP (Universal Pyrometer Protocol) · 30
temperature range:
YYYY = beginning of temp. range
ZZZZ = end of temp. range
temperature sub
range:
measuring values:
AAmsXXX
XXX = No. of measuring values
Bit 0 = 1: EEPROM error
Bit 1 = 1: Watch dog Reset
Bit 2 = 1: Under voltage reset
parameters:
Digit 1 and 2 (10 to 99 or 00): emissivity
Digit 3 (0 to 6): t
90
(exposure time)
Digit 4 (0 to 8): t
CL
(storage delete mode)
Digit 5 (0 to 1): analog output
Digits 6 and 7 (00 to 99): temperature
Digits 8 and 9 (00 to 97): address
Digit 10 (0 to 6 or 8): baud rate
Digit 11 always 0
XX: Device address decimal
00 to 97 variable device addresses
98 = global address without response (only setting commands!)
99 = global address with response
0: 1200 Bd 3: 9600 Bd 6: 57600 Bd
1: 2400 Bd 4: 19200 Bd 7: not allowed
2: 4800 Bd 5: 38400 Bd 8: 115200 Bd
temperature:
pyrometer
temperature:
Output: XXX (dec. 000 ... 099
XX = 00 to 99 relative delay value
temperature:
e.g. 0258 corresponds to 600 degrees
temperature:
XXXX = value of ambient temperature, 4
XXXX e.g. FFEC corresponds to –20 degrees
- 99dez = FF9Dhex means: automatic, no manual
compensation
ambient
temperature:
Output: marginal values for the entry, 2 x 4
e.g. FF9D0384 corresponds to -99 to 900 degrees
maximum/minimum
values:
0 = maximum value, 1 = minimum value
maximum/minimum
values:
0 = maximum value, 1 = minimum value