IS 12 AI & IS 12-AI/S Introduction 10
0.1 °C at interface and display,
< 0.025% of temperature range at the analog output
power supply, digital interface, analog output are galvanically isolated
against each other and housing
adjustable at the instrument or via serial interface:
emissivity ; response time t
90
; clear times t
CL
of the maximum value
storage; temperature sub range; analog output 0 to 20 or 4 to 20 mA;
switch points for limit contacts; °C or °F; interface RS232 or RS485; address;
baud rate; test current output
additionally (only via interface adjustable):
key lock, wait time, recalibration (with special software)
Emissivity
:
0.100 to 1.000 in 1/1000 steps
90
< 1.5 ms (with dynamical adaptation at low signal levels);
adjustable at 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s
storage:
in single or double storage. Clearing with adjusted time t
clear
0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via interface or automatically with the
next measuring object
over relay contacts), switch power max. 30 W
(I
max
: 1 A, U
max
: 60 V DC)
Rise and fall times: exposure time of the pyrometer + 2 ms
0.3% of measured value in °C + 1 °C
( = 1, t
= 1 s, T
= 15...40 °C, T
400 °C)
0.1% of measured value in °C + 1 °C
(nur IS 12-Al/S):
scanning frequency: 4 to 10 Hz
4 keys, operate with tip of ball
non condensing conditions
optionally with an additional built-in laser targeting light
(max. power level < 1 mW, = 630-680 nm, CDRH class II)
according to EU directives about electromagnetic immunity
Note: The calibration / adjustment of the instruments was carried out in accordance
with VDI/VDE directive “Temperature measurement in industry, Radiation
thermometry, Calibration of radiation thermometers”, VDI/VDE 3511, Part 4.4.
For additional details on this directive, see http://info.lumasenseinc.com/calibration or
order the directive from “Beuth Verlag GmbH” in D-10772 Berlin, Germany.