IS 6 Advanced / IGA 6 Advanced Manual Introduction · 9
2 Introduction
2.1 Appropriate use
The IMPAC IS 6 Advanced / IGA 6 Advanced are short wave infrared temperature measuring
devices with digital signal processing. They are used for non-contact temperature measurements
on metals, ceramics, graphite, etc. with a temperature range between 250 and 3000 °C.
2.2 Scope of delivery
Pyrometer, PC adjustment, and evaluation software InfraWin, works certificate, and operating
instructions.
Note: A connection cable is not included with the instrument and has to be ordered
separately (see Chapter 8, Reference numbers).
2.3 Technical Data
Measurement
Temperature Ranges:
250 to 1800 °C (MB 18), IGA
600 to 3000 °C (MB 30), IS
600 to 1800 °C (MB 18), IS
Any range adjustable within the tem
perature range, minimum span
50 °C
IGA: 1.45 to 1.8 µm
0.1 °C or 0.2 °F at interface;
< 0.0015% of adjusted temperature range
at analog output, 16 bit; 1 °C or 1 °F on display
90
120 microseconds (for IGA 6, it is recommended for measu
temperatures below 300 °C to set a response time of 1 ms (min);
adjustable to min; 1 ms; 3 ms; 5 ms; 10 ms; 50 ms; 250 ms; 1 s; 3 s; 10 s
(e = 1, t
= 1 s, T
= 25 °C)
< 1500 °C: 0.3% of reading in °C + 2 °C
> 1500 °C: 0.6% of reading in °C
(e = 1, t
= 1 s, T
= 25 °C)
Optics
Sighting: Built-in laser aiming light (max. power level < 1 mW, λ = 630 to 680 nm,
CDRH class II) or through-lens sighting
Manually focusable from rear cover measuring distance
a = 210 to 5000 mm
IS: approx. 350 : 1
Comentarios a estos manuales