IMPAC pyrometers ISR 12-LO/GS
21
9.6 Number of devices
With a click on “Number of devices” InfraWin changes to the display of 1 or 2 devices. If 2 devices
are selected, always 2 windows are displayed for settings or evaluation.
9.7 Basic settings
Under pyrometer parameters all preset values
can be displayed and modified if necessary. If the
window pyrometer parameters displayed, chang-
ing of settings on the pyrometer is blocked.
The window pyrometer parameter contains all parameter
settings described in section 8, Parame
ters.
Notes:
“Basic range” displays the total range of the pyrometer
automatically and cannot be changed. If the sub range
is changed, confirm it with the small OK button which
appears.
Under “Material “ you have the possibility to store the
names of different measuring objects with their emis-
sivity values and to recall them from the list.
A click on the laser targeting light icon (
) turns the
laser targeting light on or off at this point. After approx.
two minutes the laser targeting light is switched off
automatically.
The open / save button enable to store and recall own pyrometer configurations.
“1 meas.” shows the last pouring stream temperature in the pyrometer parameters window for
approx. 1 second.
If the pyrometer is used in another mode than pouring stream, „1 meas.“ shows the current
measuring temperature for approx. 1 second.
In this window the pouring stream mode can
be switched on and off (standard setting at the
ISR 12-LO/GS is “on”).
Descriptions to the parameters are to find un-
der 8.1 pouring stream mode).