IMPAC pyrometers ISR 12-LO/GS
27
98 = global address without response (only setting commands!)
Baud rate: AAbrX X = 1 ... 6 or 8 (dec.)
1 = 2400 baud 4 = 19200 baud (7 is not available)
2 = 4800 baud 5 = 38400 baud 8 = 115200 baud
3 = 9600 baud 6 = 57600 baud
Temp. display °C / °F AAfhX Output: X = 0: display in °C; X = 1: display in °F
Reading signal strength AAtr Output: XXXX, 1000 equivalent to 100.0%
Wait time: AAtwXX XX = 00 ... 99 (decimal)
Warning level
dirty window
AAdwXX XX = 00 … 99; 2 digit, dec. in %
Error status: AAfs Output: XX;
XX=00…FF (00 = no error)
(01…FF: error code for LumaSense service)
Internal temperature: AAgt
Output: XXX (dec. 000 ... 099, at °C, 032 ... 210 at °F)
Max. internal temp.: AAtm Output: XXX (dec. 000 ... 099, at °C, 032 ... 210 at °F)
Reading interface type: AAin
Output: X = 1 or 2; 1 = RS232, 2 = RS485
Laser targeting light: AAlaX
X = 0 switch off laser targeting light
X = 1 switch on laser targeting light
Lock keyboard: AAlkX
X = 0 ... 3
1 = lock lk1, removal with command lk0 or power off-on
0 = removal of lock lk1
3 = continuous lock lk3, removal only with command lk2
2 = removal of lock lk3
Reading parameters: AApa Output decimal 11-digit:
Digit 1 und 2 (10...99 or 00): ε
Digit 3 (0 ... 6): t
90
(exposition time)
Digit 4 (0 ... 8): t
CL
(max. storage clear mode)
Digit 5 (0 ... 1): analog output
Digit 6 und 7: (00 ... 98): temperature
Digit 8 und 9 (00 ... 97): address
Digit 10 (0 ... 6 or 8): baud rate
Digit 11 (0): (0 or 1): keyboard active or locked
Digit 12 ... 15 (0800 ... 1200): emissivity slope
Autoreset: AAre Activating auto-reset, initializes the pyrometer new and
resets the instrument to default settings
Reference number: AAbn Output: XXXXXX (hex 6-digit)
Serial number: AAsn Output: XXXX (hex 4-digit)
Device type /
software version:
AAve Output: XXYYZZ (6-digit decimal)
XX = 06 (ISR 12-LO)
YY = month of software version
ZZ = year of software version
Software version
in detail:
AAvs tt.mm.yy XX.YY
tt = day; mm = month; yy = year; XX.YY = software version
Device type: AAna Output: “ISR 12-LO“ (16 ASCII-characters)
Special pouring stream commands
Description Command Parameters
Reading of data AAtg Output: ABBBCCCC
A = serial no. of pouring date, Ascii-Hex 0 … F
BBB = time of pouring Ascii-Hex 000 … FFF in
1
/
10
s
(pre-run time + measuring time)
CCCC = temp. value Ascii-Hex 0000 … FFFF in
1
/
10
°C or °F
Pouring stream mode AAtoXX Output: XX XX = 00 pouring stream mode off
XX = 01 pouring stream mode on
Measuring condition AAtqXX XX = 00 … 99; 2 digit, dec. in % of 1024 meas. values
Pre-run time AAtpXX XX = 00 … 99; 2 digit, dec. in
1
/
10
s
Measuring time AAmtXX XX = 00 … 99; 2 digit, dec. in
1
/
10
s; 00=Automatic mode
Follow-up time AAtnXX XX = 00 … 99; 2 digit, dec. in
1
/
10
s
Note: the letter “l” means the lower case letter of “L”.